:CALCulate:TRACe{Tr}:TIME:IMPulse:WIDTh

Type

Command/Query

Syntax

:CALCulate:TRACe{Tr}:TIME:IMPulse:WIDTh <numeric>

:CALCulate:TRACe{Tr}:TIME:IMPulse:WIDTh?

Description

This command sets impulse width value for the transform function.

Parameter

Description

Transform function impulse width value

Data Type

Double precision floating point type (Double)

Range

30.2p to 20.8n

Preset Value

35p

Unit

sec

Resolution

-

Query Response

{numeric}<newline><^END>

Examples

10 OUTPUT 717;":CALC:TRAC1:TIME:IMP:WIDT 17e-12"
20 OUTPUT 717;":CALC:TRAC1:TIME:IMP:WIDT?"
30 ENTER 717;A

Related Commands

:CALCulate:TRACe{Tr}:TIME:STIMulus

Equivalent GUI

TDR/TDT > Parameters > Lowpass Impulse (In Stimulus)